Choosing the Correct Temperature Inserts

Choosing the Correct Temperature Inserts
Choosing the Correct Temperature Inserts

 

Using inserts from other calibrators
Often customers believe that they can use inserts from other models of temperature calibrators. Even if the diameter is right, the length might not be. In this case the insert is too tall. This gives a heat flow out of the top of the well and the temperature uniformity will change dramatically.


 

Using inserts from other calibrators
Using inserts from other calibrators

Using inserts that are too small 
Using inserts with diameters that are too small creates another problem. The heat transfer between the well and the insert is insufficient, as the air around the tube has an insulating effect. This will make the temperature very unstable and unable to reach the set value.


This problem also applies to the use of 6 mm and 1/4 in (6.35 mm) sensor holes. Even though this combination may seem to match it gives a very poor heat transfer as the sensors do not have an even contact with the insert.

 

Using inserts that are too small 
Using inserts that are too small 


Using inserts made of the “wrong” material   
Home-made inserts can be made in a very professional way with the correct length and size, but the material used for the inserts might cause problems and wrong results. If the specifications e.g. apply to a brass insert, it is not a good idea to use inserts made of aluminum or other materials due to the difference in conductivity.


It is therefore extremely important to use the right material for the inserts.

 

Using inserts made of the “wrong” material
Using inserts made of the “wrong” material


Using inserts with too short sensor holes   
When drilling your own inserts it is a common mistake to make the sensor holes too short. Using this kind of insert may influence the temperature gradient, as the calibrator was originally calibrated at the bottom of the well.

 

Using inserts with too short sensor holes  
Using inserts with too short sensor holes  


Using inserts without using an insulation plug    

All our cooling temperature calibrators are delivered with an insulation plug, to be placed on top of the well. The design requires this device, and the calibrators have also been calibrated with the insulation plug.


There is a large heat flow out of the well, and if the plug is not used during calibration, it will cause problems reaching the set temperature of the calibrator.

 

Using inserts without using an insulation plug   
Choosing the Correct Temperature Inserts

 

Using inserts with the correct size, holes and material
To avoid problems, waste of time and to get the optimum use of your calibrator we recommend that you use inserts from AMETEK Test & Calibration Instruments as they follow our specifications.